Coating Thickness
Call Me Free
Fischer measurement Technologies (India) Pvt. Ltd.

FISCHERSCOPE X-RAY XDL Series XDL

FISCHERSCOPE X-RAY XDL X-ray fluorescence measuring device for manual as well as automated coating thickness measurements on decorative and protective coatings. They are universally applicable energy dispersive X-ray measuring instruments used for non-destructive thickness measurements as well as analysis of thin coatings for the measurements on mass-produced parts as well as printed circuit boards. FISCHERSCOPE X-RAY XDL is designed under the direction of practiced quality controllers following the international industrial standards. 

Models

  • XDL 210: Plane support stage, fixed Z-axis

  • XDL 220: Plane support stage, motor-driven Z-axis

  • XDL 230: Manually operable XY-stage, motor-driven Z-axis

  • XDL 240: Motor-driven XY-stage that moves into the loading position automatically, when the protective hood is opened. Motor-driven programmable Z-axis

 

Applications

  • Measurement of electroplated mass-produced parts

  • Study of thin coatings like decorative chromium-plating

  • Examination of functional coatings in the electronics as well as semiconductor industries

  • Automated measurements such as printed circuit boards

  • Solution study in the electroplating

 

General Specification

  • Intended use:Energy dispersive x-ray fluorescence measuring device (EDXRF) to find out thin coatings and solution study.
  • Element range:Chlorine (17) to Uranium (92) - up to 24 elements simultaneously with option WinFTM® BASIC.
  • Design:Bench top unit with upwards opening hood
  • Measuring direction:Top down

X-Ray Source

  • X-ray tube:Tungsten tube with beryllium window
  • High voltage:Three steps: 30 kV, 40 kV, 50 kV
  • Aperture (Collimator): 0.3 mm (11.8 mils), Optional:  0.1 mm (3.9 mils), 0.2 mm (7.9 mils), slot 0.3 x 0.05 mm (11.8 x 2 mils)
  • Measurement spot: On the basis of measuring distance and on the aperture, the real measurement. Spot size is shown in the video image.
  • Smallest measurement spot: approx. 0.2 mm (7.9 mils)

X-Ray Detection

  • X-ray detector:Proportional counter tube
  • Measuring distance:0 - 80 mm (0 - 3.2 in)
  • Distance compensation with patented DCM technique for simplified measurements at changeable distances. For different applications or for higher needs on accuracy an extra calibration might be needed.

Sample Alignment

  • Sample positioning:Manually
  • Video microscope:High-resolution CCD color camera is designed for optical monitoring of the position along the Cross hairs with a calibrated scale (ruler), spot-indicator, Adjustable LED illumination and primary beam axis.
  • Zoom factor:Digital 1x, 2x, 3x, 4x

Evaluation Unit

  • Computer:Windows®-PC
  • Software:Standard: Fischer WinFTM® LIGHT

Optional: Fischer WinFTM® BASIC, PDM®, SUPER

Sample support stage

Sample Stage


XDL 210

XDL 220

XDL 230

XDL 240

Design

Fixed sample Support

Fixed sample Support

Manual XY-stage

Programmable XY-stage

Maximum travel XY

-

-

95 x 150 mm (3.7 x 5.9 in)

255 x 235 mm (10 x 9.2 in)

Travel speed XY

-

-

-

80 mm/s (3.1 in/s)

Repeatability precision XY

-

-

-

0,01 mm (0.4 mils)(*1)

Usable sample placement area

463 x 500 mm (18.2 x 19.7 in)

463 x 500 mm (18.2 x 19.7 in)

420 x 450 mm (16.5 x 17.7 in)

300 x 350 mm (11.8 x 13.8 in)

Z axis

Fixed Position (Top/Middle/Bottom)

Electrically adjustable

Electrically adjustable

Programmable

Travel Z axis

-

140 mm (5.5 in)

140 mm (5.5 in)

140 mm (5.5 in)

Max. sample weight

20 kg (44 lb)

20 kg (44 lb)

20 kg (44 lb)

5 kg (11 lb) / 20 kg (44 lb)(*2)

Max. sample height

155/90/25 mm (6/3.5/1 in)

140 mm (5.5 in)

140 mm (5.5 in)

140 mm (5.5 in)

Laser pointer (class 1) to
support accurate sample placement

-

-

Yes

Yes

 

Electrical Data

  • Power supply:AC 115 V or AC 230 V 50 / 60 Hz
  • Power consumption:max. 120 W, without evaluation PC
  • Protection class:IP40

Dimensions

  • External dimensions:Width x depth x height [mm]: 570 x 760 x 650 mm, [in]: 22 x 30 x 26
  • Interior dimensions:Width x depth x height [mm]: 460 x 495 x see max. sample height measurement:
  • chamber:[in]: 18 x 19.5 x 26 x see max. sample height
  • Weight:XDL 210: 94 kg (207 lb); XDL 220: 99 kg (218 lb); XDL 230: 107 kg (235 lb);XDL 240: 120 kg (264 lb)

Environmental Conditions

  • Operating temperature:10oC- 40oC / 50oF o104oF
  • Storage/Transport 0oC - 50oC / 32oF - 122oF temperature:
  • Admissible air humidity:  95 %, non-condensing

Standards

  • CE approval:EN 61010
  • X-Ray standards:DIN ISO 3497 and ASTM B 568
  • Approval:Fully protected device with type approval according to the German regulations.

 Product details  

Repeatability

Greater than 1 % for gold,measurement time 60 sec

Brand

Fischer

Power Type

Electric

Operation Mode

Automatic

Dimension

403 x 588 x 365 mm

Power Supply

AC 230 V 50 / 60 Hz