XRF Spectrometer for PCB XDLM PCB 210
PCB Coating Thickness Measurement Gauge
Coating thickness measurement for PCB using XDLM PCB 210 is a qualitative range of devices available in various technical specifications. These are extensively acknowledged as specific X-ray fluorescence measuring instruments for measurements and analyses of coating thicknesses and compositions on printed circuit boards. These are acclaimed for its longer service life and outstanding accuracy.
Applications
- Measurements on small components and structures on printed circuit boards in sizes up to 610 x 610 mm (24 x 24 in)
- Measurements of functional coatings in the electronics and semiconductor industries
- XDLM-PCB 210 and 220: Automated measurements, e.g., in quality control
- Determining the composition of electroplating baths
General Specification
- Intended use: Energy dispersive x-ray fluorescence measuring instrument (EDXRF) to determine thin coatings, small structures and alloys.
- Element range: Chlorine (17) to Uranium U (92) up to 24 elements simultaneously.
- Design: Bench-top unit with housing with a slot on the side
- Measuring direction: Top down
X-Ray Source
X-ray tube:
- High voltage: Three steps: 30 kV, 40 kV, 50 kV
- Aperture (Collimator):XDLM-PCB 200/210: 0.1 mm (3.9 mils), optional 0.2 mm (7.9 mils), slot 0.3 x 0.05 mm (11.8 x 2 mils)
- Standard (523-440): 0.1 mm (3.9 mils); 0.2 mm (7.9 mils); 0.05 x 0,05 mm (2 x 2 mils); 0.2 x 0.03 mm (7.9 x 1.2 mils)
- Optional (523-366): 0.1 mm (3.9 mils);0.2 mm (7.9 mils); 0.3 mm (11.8 mils); 0.3 x 0.05 mm (11.8 x 2 mils)
- Measurement spot: Depending on the measuring distance and on the aperture, the actual measurement. Spot size is shown in the video image.
- Smallest measurement spot: Approx. 0.2 mm (7.9 mils)
- Measurement distance: 0-10 mm (0-0.4 in) Distance compensation with patented DCM for simplified measurements at varying distances. For particular applications or for higher demands on accuracy an additional calibration might be necessary.
X-Ray Detection
- X-ray detector:Proportional counter tube
Sample Alignment
- Sample positioning:Manually
- Video microscope:High-resolution CCD color camera for optical monitoring of the measurement location along the primary beam axis, Manual focusing and auto-focus, Crosshairs with a calibrated scale (ruler) and spot-indicator, Adjustable LED illumination,Laser pointer (class 1) to support accurate,specimen placement.
- Zoom factor:Digital 1x, 2x, 3x, 4x
Evaluation Unit
- Computer:Windows-PC
- Software:Standard: Fischer WinFTM LIGHT
- Optional: Fischer WinFTM BASIC, PDM̨̉, SUPER
Dimensions
Dimensions |
| XDLM-PCB 200 | XDLM-PCB 210 |
External dimensions | 610 x 750 x 450 mm | With maximum XY travel range: 1000 x 1265 x 470 mm (39.4 x 49.8 x 18.5 in) |
Width x depth x height | With extension: 1200 x 1050 x 450 mm (47.2 x 41.3 x 17.7 in) | XY table retracted in home position: 650 x 810 x 470 mm (25.6 x 31.9 x 18.5 in) |
Weight | Approx. 86 kg (190 lb) | Approx. 86 kg (190 lb) |
Sample support stage
Sample Stage |
| XDLM-PCB 200 | XDLM-PCB 210/220 |
Design | Fixed sample support with manual pop out function | Programmable XY-stage with pop out function |
Usable sample placement area | 600 x 600 mm (23.6 x 23.6 in) With extension: 1200 x 900 mm (47.2 x 35.4 in) | 600 x 600 mm (23.6 x 23.6 in) |
Maximum travel XY-axis | - | 450 x 300 mm (17.7 x 11.8 in) |
Max. travel speed XY | - | 450 x 300 mm (17.7 x 11.8 in) 60 mm/s (2.4 in/s) |
Repeatability precision XY | - | âââ¬Â°Ã¤ 0.01 mm (0.4 mils), unidirectional |
Max. sample Weight | 5 kg (11 lb) | 5 kg (11 lb) |
Max. sample Height | 5 mm | 5 mm |
Electrical Data
- Power supply:AC 115 V or AC 230 V 50 / 60 Hz
- Power consumption:max. 120 W
- Protection class:IP40
Environmental Conditions
- Operating temperature:10 - 40°C / 50- 104 °F
- Storage/Transport0- 50 °C / 32 °F 122 °F temperature:
- Admissible air humidity:95 %, non-condensing
Standards
- CE approval:EN 61010
- X-Ray standards:DIN ISO 3497 and ASTM B 568
- Approval:Fully protected instrument with type approval according to the German regulations.
Product details
Protection Class | IP40 |
Brand | Fischer |
Dimension | 403 x 588 x 444 mm |
Weight | Approx 45 kg |
Power Supply | AC 115 V or AC 230 V 50 / 60 HzAC 115 V or AC 230 V 50 / 60 Hz |
Application/Usage | Industrial |